Skip to content
MRI Academy

The Spin-Echo Family

SE, FSE/TSE, inversion recovery, FLAIR & STIR

Core⏱ ~55 minPulse-sequence diagramContrast playground

The 180° refocusing pulse recovers signal lost to static dephasing and defines the most robust contrast in MRI. We build spin echo, accelerate it into fast/turbo spin echo with echo trains, and add inversion pulses to null fat (STIR) or CSF (FLAIR).

By the end you will be able to

  • 1Explain how a 180° pulse refocuses static dephasing to form a spin echo
  • 2Contrast SE with FSE/TSE echo trains and the meaning of the echo-train length
  • 3Compute the inversion time needed to null a tissue of known T1
  • 4Select STIR vs FLAIR vs chemical fat saturation for a clinical goal

Prerequisites: T1, T2 & T2*: Relaxation Mechanisms, k-Space & the Fourier Transform

Why refocus at all?

After a single 90 degree excitation, the transverse magnetization does not simply decay at the rate set by spin-spin interactions. It decays much faster, at the rate 1/T21/T_2^{*}, because spins at different positions precess at slightly different frequencies. Some of this spread is random and irreversible (true T2T_2 from fluctuating local fields and diffusion), but much of it is static and reversible: fixed magnetic-field inhomogeneity from an imperfect magnet, susceptibility gradients at tissue interfaces, and chemical-shift offsets. The genius of the spin echo, introduced by Erwin Hahn in 1950, is that a single 180180^{\circ} refocusing pulse can undo all of the static dephasing and recover a signal that reflects true T2T_2.

This single idea — turn an irreversible-looking decay into a refocusable one — is the foundation of an entire family of sequences: the basic spin echo (SE), fast or turbo spin echo (FSE/TSE), and inversion-recovery variants including STIR and FLAIR. This lesson derives the echo, then builds the family on top of it.

Deriving the Hahn echo

Consider an isochromat (a small group of spins) at position experiencing a constant frequency offset Δω\Delta\omega relative to the rotating-frame reference. After the 9090^{\circ} pulse places magnetization in the transverse plane, the accumulated phase at time tt is simply ϕ(t)=Δωt\phi(t) = \Delta\omega \, t. Different isochromats have different Δω\Delta\omega, so the phases fan out and the vector sum (the measurable signal) collapses on the timescale T2T_2^{*}.

Now apply a 180180^{\circ} pulse at time τ\tau. A refocusing pulse about the yy-axis maps phase ϕπϕ\phi \rightarrow \pi - \phi, which is equivalent to negating the accumulated phase. Just after the pulse the isochromat phase is Δωτ-\Delta\omega\,\tau. Precession continues at the same Δω\Delta\omega, so at a later time tt the phase is:

Eq. 5.1
Phase after a 180 degree pulse applied at time tau
ϕ(t)=Δωτ+Δω(tτ)=Δω(t2τ)\phi(t) = -\Delta\omega\,\tau + \Delta\omega\,(t-\tau) = \Delta\omega\,(t - 2\tau)
ϕ\phi
accumulated phase of an isochromat in the rotating frame
Δω\Delta\omega
static frequency offset of the isochromat
τ\tau
time between the 90 degree and 180 degree pulses

The phase becomes exactly zero when t=2τt = 2\tau, independent of Δω\Delta\omega. Every isochromat, no matter how far off-resonance, returns to zero phase at that instant. The fan of vectors collapses back to alignment and the signal rephases into an echo. The echo time is therefore TE=2τTE = 2\tau, and the static dephasing has been completely undone.

What the refocusing cannot undo is the random component. Molecular tumbling causes the local field — and hence Δω\Delta\omega — to fluctuate during the experiment, so each spin's phase history is not perfectly retraced. This irreversible loss is true T2T_2 decay. The amplitude of the echo is therefore attenuated only by T2T_2, not T2T_2^{*}:

Eq. 5.2
Spin-echo signal at the echo peak
S(TE)=S0eTE/T2S(TE) = S_0 \, e^{-TE/T_2}
S0S_0
signal immediately after excitation (proportional to proton density and T1 recovery)
TETE
echo time, equal to twice the 90 to 180 spacing
T2T_2
true transverse relaxation time of the tissue

Multiple echoes and the CPMG train

Because the 180180^{\circ} pulse only inverts phase, you can apply a series of them after one excitation and collect an echo after each. With refocusing pulses spaced evenly, echoes form at TE=2τ,4τ,6τ,TE = 2\tau, 4\tau, 6\tau, \dots, and their peak amplitudes trace out the T2T_2 decay curve. Fitting lnS\ln S versus TETE to a straight line yields T2T_2 directly — the classic multi-echo measurement.

Two practical problems arise. First, an imperfect 180180^{\circ} pulse (a true flip angle of, say, 170170^{\circ}) leaves residual longitudinal magnetization and accumulates error over many echoes. Second, diffusion through residual gradients adds spurious decay. The Carr-Purcell-Meiboom-Gill (CPMG) scheme solves both: the 180180^{\circ} pulses are phase-shifted by 9090^{\circ} relative to the excitation (applied about yy while excitation is about xx), so flip-angle errors on even echoes cancel those on odd echoes. CPMG is the workhorse refocusing scheme used in essentially every modern spin-echo train.

Fast (turbo) spin echo: encoding many lines per TR

Conventional SE acquires one phase-encoding line per TR, so a 256256-line scan with TR=4000TR = 4000 ms takes over 1717 minutes. Fast spin echo (FSE on GE, TSE on Siemens and Philips) collapses this by phase-encoding each echo of a CPMG train separately. Each 180180^{\circ} pulse produces an echo that gets its own phase-encode gradient (applied before readout and rewound after), so a single TR fills several lines of k-space at once.

The number of echoes used per TR is the echo-train length (ETL), also called the turbo factor. The acceleration is exactly the turbo factor: a turbo factor of 1616 on a 256256-line acquisition needs only 256/16=16256/16 = 16 TRs, cutting a 1717-minute scan to about a minute. The price is that the different lines of k-space are acquired at different real echo times along the decay curve.

Effective TE and k-space ordering

Because contrast in an image is dominated by the center of k-space (low spatial frequencies carry the bulk of the signal energy and gross contrast), the apparent contrast of an FSE image is set by whichever echoes are mapped to the central lines. That echo time is the effective TE (TEeffTE_{eff}). To produce a T2-weighted image, the sequence orders the train so that late echoes (long real TE) land in the center of k-space; for a proton-density image, early echoes fill the center.

| Property | Conventional SE | Fast / turbo SE | | --- | --- | --- | | Lines per TR | 1 | Equal to the turbo factor (e.g. 4 to 32) | | Scan time for 256 lines | 256 times TR | 256 / turbo factor times TR | | Contrast set by | Single chosen TE | Effective TE from central k-space echo | | Edge sharpness | Uniform | Slight blur from T2 decay across the train | | Fat signal on T2 | Moderate | Brighter (J-coupling decoupled) | | SAR | Lower | Higher (many 180 degree pulses) |

FSE is not free of artifact. Because outer k-space lines are filled by echoes that have decayed further, the high-spatial-frequency data are attenuated, which blurs edges along the phase-encode direction. Longer turbo factors and tissues with short T2T_2 worsen this. A second hallmark is bright fat: in conventional SE, J-coupling among the protons of fatty-acid chains causes additional dephasing that darkens fat on T2 images, but the rapid train of 180180^{\circ} pulses in FSE disrupts (decouples) this J-modulation, so fat stays conspicuously bright. This is why a separate fat-suppression technique is usually paired with T2 FSE.

Inversion recovery, STIR, and FLAIR

Inversion recovery (IR) prepends a 180180^{\circ} inversion pulse before the spin-echo excitation. The inversion flips MzM_z from +M0+M_0 to M0-M_0, and during the inversion time (TITI) the longitudinal magnetization recovers along its T1T_1 curve, passing through zero on its way back to equilibrium:

Eq. 5.3
Longitudinal recovery after a 180 degree inversion
Mz(TI)=M0(12eTI/T1)M_z(TI) = M_0\left(1 - 2\,e^{-TI/T_1}\right)
Mz(TI)M_z(TI)
longitudinal magnetization at the time of excitation
M0M_0
equilibrium magnetization
TITI
inversion time between the 180 degree inversion and the 90 degree excitation
T1T_1
longitudinal relaxation time of the tissue

A tissue is nulled when Mz(TI)=0M_z(TI) = 0, which occurs when 12eTI/T1=01 - 2e^{-TI/T_1} = 0, giving the famous null condition:

Eq. 5.4
Inversion time that nulls a given tissue
TInull=T1ln20.693T1TI_{null} = T_1 \, \ln 2 \approx 0.693 \, T_1
TInullTI_{null}
inversion time at which a tissue gives zero signal
T1T_1
longitudinal relaxation time of the tissue to be nulled

Choosing TITI to null a specific tissue is the entire point of STIR and FLAIR. STIR (Short TI Inversion Recovery) nulls fat: at 1.51.5 T fat has T1250T_1 \approx 250 ms, so TInull0.693×250170TI_{null} \approx 0.693 \times 250 \approx 170 ms (typically set around 150150 to 170170 ms). Because STIR suppresses fat by its short T1T_1 rather than by chemical shift, it is robust to field inhomogeneity and works well over large fields of view and near metal. FLAIR (Fluid-Attenuated Inversion Recovery) nulls CSF: CSF has T14000T_1 \approx 4000 ms, so TInull0.693×40002500TI_{null} \approx 0.693 \times 4000 \approx 2500 to 28002800 ms. FLAIR is the standard for seeing periventricular and cortical lesions that would otherwise be hidden by bright CSF on T2.

| Sequence | Tissue nulled | Approx T1 at 1.5 T | Approx TI | | --- | --- | --- | --- | | STIR | Fat | 250 ms | 150 to 170 ms | | FLAIR | CSF (free water) | 4000 ms | 2200 to 2800 ms | | IR (T1 weighting) | None (used for high T1 contrast) | varies | 300 to 700 ms |

How TR and TE set the weighting

Spin-echo image contrast is governed by two operator-chosen parameters. TR controls how much T1T_1 recovery occurs between excitations and therefore the T1T_1 weighting; TE controls how much T2T_2 decay occurs before sampling and therefore the T2T_2 weighting. Combining the recovery and decay terms gives the working signal equation for spin echo:

Eq. 5.5
Spin-echo signal as a function of TR and TE
Sρ(1eTR/T1)eTE/T2S \propto \rho \left(1 - e^{-TR/T_1}\right) e^{-TE/T_2}
ρ\rho
proton (spin) density
TRTR
repetition time between successive excitations
TETE
echo time
T1,T2T_1, T_2
tissue relaxation times
  • T1-weighted: short TR (around 400 to 700 ms) maximizes T1 differences, short TE (around 10 to 20 ms) minimizes T2 influence. Fat bright, fluid dark.
  • T2-weighted: long TR (2500 ms or more) removes T1 dependence, long TE (around 80 to 120 ms) emphasizes T2 differences. Fluid and many lesions bright.
  • Proton-density (PD): long TR removes T1 dependence, short TE removes T2 dependence, leaving contrast driven by spin density.
  • Avoid: short TR with long TE produces mixed weighting and low signal — generally not diagnostically useful.

The interactive diagram below lets you place the 9090^{\circ} excitation, the 180180^{\circ} refocusing pulse, and the readout, and watch where the echo forms relative to the pulse spacing. Set up a single-echo SE first, then add refocusing pulses to build a CPMG/FSE train and observe how the echoes line up at multiples of 2τ2\tau.

The 180° refocusing pulse at TE/2 reverses static dephasing, producing a spin echo at TE that is immune to fixed field inhomogeneity (true T2).

Now use the contrast playground to sweep TRTR and TETE across tissues with different T1T_1 and T2T_2. Try to reproduce the three canonical weightings from Eq. 5.5: drive TR short for T1 weighting, push TE long for T2 weighting, and find the long-TR short-TE corner that isolates proton density.

T1-weighted

Short TR + short TE

Sequence
Field
S ∝ PD · (1 − e^(−TR/T1)) · e^(−TE/T2)
White matter
0.36
Gray matter
0.26
CSF
0.13
Fat
0.76
Lesion (edema)
0.29

SAR: the cost of all those 180s

Every RF pulse deposits energy in tissue, quantified as the specific absorption rate (SAR) in watts per kilogram. The energy of a pulse scales roughly with the square of the flip angle and, critically, with the square of the main field strength B0B_0, because the Larmor frequency — and thus the RF frequency — rises with B0B_0. A 180180^{\circ} refocusing pulse deposits about four times the energy of a 9090^{\circ} pulse of the same shape.

Eq. 5.6
Approximate SAR scaling for a refocused sequence
SARB02α2N180TR\mathrm{SAR} \propto B_0^{2} \, \alpha^{2} \, \frac{N_{180}}{TR}
B0B_0
main magnetic field strength
α\alpha
refocusing flip angle
N180N_{180}
number of 180 degree pulses per TR (the turbo factor)
TRTR
repetition time

FSE multiplies the number of 180180^{\circ} pulses per TR by the turbo factor, so a long echo train is an RF-heavy sequence. Moving from 1.51.5 T to 33 T roughly quadruples SAR for the same pulses. This is the dominant constraint on high-field FSE: scans can hit the regulatory SAR limit and the scanner must lengthen TR, drop slices, or reduce the refocusing flip angle. The standard mitigation is variable-flip-angle (hyperecho or TRAPS-style) refocusing, where the 180180^{\circ} pulses are replaced by lower flip angles (often 120120^{\circ} or less); this dramatically cuts SAR while a carefully designed flip-angle sweep maintains usable signal across a very long train (the basis of single-slab 3D FSE sequences such as SPACE, CUBE, and VISTA).

Imaging for this lesson

Explore the correct real MRI for this topic — yours to scroll, window and render.

Spin-echo contrasts on the brain — and the cervical spine and neck, where T2 spin-echo is the everyday workhorse. The 180° refocusing pulse is what makes these robust true-T2 images possible.

Brain & head

Spine & neck

View
Colormap

Scroll to change slice · click-drag to move the crosshair · right-click-drag to window (brightness/contrast).

Check your understanding

  1. 1.In a Hahn spin echo with the 180 degree refocusing pulse applied at time tau after excitation, at what time does the echo form?

  2. 2.Compared with a gradient echo, the spin-echo signal at the echo peak is attenuated according to which decay constant?

  3. 3.In fast (turbo) spin echo, what primarily determines the effective TE and therefore the apparent image contrast?

  4. 4.A radiologist wants to null cerebrospinal fluid, which has a T1 of about 4000 ms at 1.5 T. Which inversion time is appropriate?

  5. 5.Why does moving a fast spin-echo sequence from 1.5 T to 3 T sharply increase the risk of exceeding SAR limits?

Keep exploring

Take this topic further on these trusted, free references:

Further reading

  • [1]Hahn EL. Spin Echoes. Physical Review. 1950;80(4):580-594.
  • [2]Bernstein MA, King KF, Zhou XJ. Handbook of MRI Pulse Sequences. Elsevier Academic Press; 2004.
  • [3]Haacke EM, Brown RW, Thompson MR, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design. Wiley; 1999.
  • [4]McRobbie DW, Moore EA, Graves MJ, Prince MR. MRI from Picture to Proton. 3rd ed. Cambridge University Press; 2017.
  • [5]Bushberg JT, Seibert JA, Leidholdt EM, Boone JM. The Essential Physics of Medical Imaging. 4th ed. Wolters Kluwer; 2020.
  • [6]Hennig J, Nauerth A, Friedburg H. RARE imaging: a fast imaging method for clinical MR. Magnetic Resonance in Medicine. 1986;3(6):823-833.